Electrolysis of Copper in Copper Sulphate Solution
|
Length: 1485 words (4.2 double-spaced pages) Rating: Red (FREE) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Electrolysis of Copper in Copper Sulphate Solution
Introduction Aim This is an investigation into how temperature affects the amount of copper lost from the Anode and gained on the Cathode during electrolysis. Electrolysis is… The chemical change by passing electric charge through certain conducting liquids (electrolytes). The current is conducted by migration of ions - negative ones from the Anode (positive electrode), and positive ones to the Cathode (negative electrode). Reactions take place at the electrodes by transfer of electrons in the direction of the current. In this case, copper electrodes in a copper sulphate solution are used. In the electrolysis of copper, copper atoms in the Anode become copper ions by losing electrons which go into the circuit. Cu - 2e à Cu The positive ions are attracted to the negative Cathode but only copper ions are discharged. Cu - 2e à Cu So this electrode gets coated with copper. Diagram [IMAGE] Also attached to this circuit will be a variable resistor. An ammeter, placed in the circuit, will register the flow of current: from this indication, the electric circuit can be deduced to be complete.Also, a stopwatch will be used. The 'cell' is a power pack. Variables The variables in this experiment are: · The electrodes · Resistance · The electrolyte · Temperature · Position of the electrodes (Distance between) · Time Only the size of the electrodes and times are being investigated, therefore in order for this to be a fair test, the other factors must be kept constant. The same CuSO will be used throughout, so the concentration is the same, and the same spacing between electrodes will be used. The size of the electrodes should be the same, but they will be reused, so the size will change from one experiment to another. Time intervals of 5 minutes for each experiment will be used, until it equals 35 minutes in total. Prediction I predict that the longer the electrolysis lasts, the more copper will be deposited onto the Cathode as the ions will be passed from the anode to the cathode. Analysis The concentration of the solution isn't expected to have a significant effect, since the concentration of ions do not make it any easier for the electrons to flow, nor does it affect any factors in the ionic equation of reaction, so it should not have an effect. However, at very low concentration, the resistance of water increases due to the lack of dissolved ions, and this will have an effect on the current and therefore having an effect on the mass of copper deposited indirectly. The time allowed for the electrolysis to run will obviously have an effect on the mass of copper deposited, since the deposition of copper is a continuous process, and providing there is a fixed current it happens at a fixed rate. Therefore, the longer the time, the more copper ions would be allowed to deposit and the bigger the combined mass. The mass should be directly proportional to time since the rate is fixed. Apparatus · 1 large beaker · 2 copper electrodes · 1 power pack · 1 ammeter · 1 variable resistor · 4 long wires · 1 stopwatch · 300 ml copper sulphate solution · 1 piece sandpaper · 1 bottle propanone Plan An experiment will be conducted to find out how much copper is dissociated from the anode to the Cathode in a certain time. The resistor will be monitored to make sure that the experiment is kept at the same resistance constantly, to prevent the occurrence of anomalous results. 300ml of COSO will be used every time the experiment is undertaken. Once the experiment has been set up, it will be initiated for 5 minutes, and then stopped. The electrodes will then be cleaned and the weight of each electrode will be recorded in a results table. The experiment will be repeated until there have been 35 minutes worth of results gathered. In addition, the same person will perform each experiment to keep it a fair test Results Two experiments were undertaken to ensure that the results were more reliable. Both are as follows: Time (minutes) Anode (+ve) Weight (g) Cathode (-ve) Weight (g) 0 72.89 92.54 5 72.88 92.58 10 72.88 92.70 15 72.84 92.51 20 72.78 92.60 25 72.74 92.63 30 72.68 92.68 35 72.60 92.73 Exp 1 - 92.70 92.51 An explanation will be given to this unlikely decrease in mass after the results of the first experiment have been analysed. Time (minutes) Anode (+ve) Weight (g) Cathode (-ve) Weight (g) 0 72.59 92.56 5 72.56 92.58 10 72.53 92.61 15 72.50 92.64 20 72.49 92.66 25 72.49 92.68 30 72.45 92.71 35 72.43 92.70 Exp 2 - 92.71 92.70 Again, an explanation of this very small decrease in mass will be given once the results of this experiment have been analysed. Time (Minutes) Original Mass Of Anode (grams) Final Mass Of Anode (grams) Change In Mass At Anode (grams) 5 72.89 72.88 -0.01 10 72.88 72.88 -0.00 15 72.88 72.84 -0.04 20 72.84 72.78 -0.06 25 72.78 72.74 -0.04 30 72.74 72.68 -0.06 35 72.68 72.60 -0.08 Comparison of increase and decrease in mass - Exp 1 Time (Minutes) Original Mass Of Cathode (g) Final Mass Of Cathode (g) Change In Mass At Cathode (g) 5 92.54 92.58 +0.04 10 92.58 92.70 +0.02 15 92.70 92.51 -0.19 20 92.51 92.60 +0.09 25 92.60 92.63 +0.03 30 92.63 92.68 +0.05 35 92.68 92.73 +0.05 Time (Minutes) Decrease in mass at Anode (g) Increase in mass at cathode (g) 5 -0.01 +0.04 10 -0.00 +0.02 15 -0.04 -0.19 20 -0.06 +0.09 25 -0.04 +0.03 30 -0.06 +0.05 35 -0.08 +0.05 As seen above, the figure -0.19 is totally unexpected. The only explanation is that a lot of copper was washed off when the cathode was cleaned. Comparison of increase and decrease in mass - Exp 2 Time (Minutes) Original Mass Of Anode (grams) Final Mass Of Anode (grams) Change In Mass At Anode (grams) 5 72.59 72.56 -0.03 10 72.56 72.53 -0.03 15 72.53 72.50 -0.03 20 72.50 72.49 -0.01 25 72.49 72.49 -0.00 30 72.49 72.45 -0.04 35 72.45 72.43 -0.02 Time (Minutes) Original Mass Of Cathode (g) Final Mass Of Cathode (g) Change In Mass At Cathode (g) 5 92.56 92.58 +0.02 10 92.58 92.61 +0.03 15 92.61 92.64 +0.03 20 92.64 92.66 +0.02 25 92.66 92.68 +0.02 30 92.68 92.71 +0.03 35 92.71 92.70 -0.01 Time (Minutes) Decrease in mass at Anode (g) Increase in mass at cathode (g) 5 -0.03 +0.02 10 -0.03 +0.03 15 -0.03 +0.03 20 -0.01 +0.02 25 -0.00 +0.02 30 -0.04 +0.03 35 -0.02 -0.01 The figure -0.01 is unexpected, but is only a very small decrease in mass, so it is very likely that this small amount of copper was washed off during the cleaning of the electrodes. Experiment 1 [IMAGE] Experiment 2 [IMAGE] Average results Time (Minutes) Decrease in mass at Anode (g) Increase in mass at cathode (g) 5 -0.02 +0.03 10 -0.015 +0.025 15 -0.035 -0.08 20 -0.035 +0.055 25 -0.02 +0.025 30 -0.05 +0.04 35 -0.05 +0.02 [IMAGE] After examining each graph carefully, it is clear that experiment 2 was more of a success. The ideal results would have made a graph something like this: [IMAGE] The results to make this graph would have been: Time (Minutes) Decrease in mass at Anode (g) Increase in mass at cathode (g) 5 -0.03 0.03 10 -0.04 0.04 15 -0.02 0.02 20 -0.03 0.03 25 -0.02 0.02 30 -0.04 0.04 35 -0.01 0.01 Evaluation Electrolysis is the name given to the chemical process which occurs, for example when an electric current is passed between two electrodes dipped into a liquid solution. There are many factors which could have made this experiment more of a success. For example, the variables could have been more carefully controlled. The electrodes; even after a good clean were still quite dirty and obviously had irremovable substances from previous experiments still attached to them. Also, if the electrodes had been closer together, there may have been more interaction/transference between them. Resistance; A more accurate resistor could have been used to ensure that the resistance was kept at a constant rate. The Electrolyte; The same CuSO and amount of it was used each time, so it is unlikely that this had any measurable effect. Temperature; A water bath could have been used to ensure a constant temperature. During the experiment, room temperature was used; this could have been rapidly reduced simply by somebody opening a window. Time; A more accurate stopwatch could have been used to ensure non-anomalous results. I predicted that the amount of copper lost on the Anode would be the amount of copper accumulated on the Cathode, but the results do not show this. In fact they are quite uneven. As ions dissociate from the Anode, an equal amount of ions should attach to the Cathode. In the experiment conducted, this did not happen. The experiment was not a success as the results do not show much even exchange of mass. It would have been preferable to repeat the experiment a minimum of ten times and to take the mean of these results. They would then have been plotted to render anomalous results invalid. During experiment 1, the Cathode decreased 19g in mass. As it has been previously explained, it is possible that 19g of copper was washed off when the electrode was cleaned in preparation for the next experiment. To ensure more reliable, accurate results the variables in this experiment could have been kept constant under more carefully controlled conditions. For example, the temperature could have been kept at something other than room temperature as it is not reliable. To do this, a water bath could be used. How to Cite this Page
MLA Citation:
"Electrolysis of Copper in Copper Sulphate Solution." 123HelpMe.com. 18 May 2013 <http://www.123HelpMe.com/view.asp?id=120594>. |
|
